Peptide COAs usually name two techniques: HPLC and LC-MS. They sound alike and are often confused, but they answer different questions, and a report that has only one of them leaves a gap.
| HPLC (with UV detection) | LC-MS | |
|---|---|---|
| Measures | How much light each separated component absorbs | The mass of each separated component |
| Typically used for | Purity (peak area share) and, against a standard, quantity | Identity and spotting impurities by mass |
| Strength | Well suited to relative amounts | Tells you what a peak actually is |
| Limit | A peak doesn’t prove identity; what doesn’t absorb UV isn’t seen | Not a precise measure of how much is there unless it’s calibrated for that |
In LC-MS the first half is the same kind of separation as in HPLC. The components leave the column one after another and, instead of (or as well as) being measured by UV light, they are fed into a mass spectrometer. That is why one instrument run can give both a chromatogram and a mass.
Why a COA should usually show both
Purity without identity says a lot of one substance was detected but not what it is. Identity without purity says the right molecule is present but not how much else is there. Read together they are far more informative than either alone.
m/z value for identity, not just the word “confirmed”.Next: reading the chromatogram and reading a mass-spec result.